====== IOP correction for Central Corneal Thickness ====== ^ Central Corneal Thickness (μm) ^ Correction Value (mmHg) ^ | 410 | +10 | | 415 | +10 | | 420 | +9 | | 425 | +9 | | 430 | +8 | | 435 | +8 | | 440 | +7 | | 445 | +7 | | 455 | +6 | | 465 | +6 | | 475 | +5 | | 485 | +4 | | 495 | +4 | | 505 | +3 | | 515 | +2 | | 525 | +1 | | 535 | +1 | | 545 | 0 | | 555 | -1 | | 565 | -1 | | 575 | -2 | | 585 | -3 | | 595 | -4 | | 605 | -4 | | 615 | -5 | | 625 | -6 | | 635 | -6 | | 645 | -7 | ===== References ===== [[https://onlinelibrary.wiley.com/doi/abs/10.1111/j.1755-3768.1975.tb01135.x|Ehlers N., Bramsen T., Sperling S. Applanation tonometry and central corneal thickness. Acta Ophthalmol. 1975;53:34–43]]\\ {{tag>glaucoma cornea}}